Parameter Robustness Analysis for System Function Reconstruction
Main Article Content
Abstract
The quality of images in system function reconstruction (SFR) depends on an extensive calibration scan that acquires the system matrix (SM). A change in parameters of the scanner, trajectory, or magnetic nanoparticle requires the acquisition of a new SM. In this work, we analyze the parameter robustness of SFR with simulations. We investigate the effect of utilizing an existing SM in the case of a change in the aforementioned parameters. The results show that a new calibration scan is not needed, as long as the change is sufficiently small.
Int. J. Mag. Part. Imag. 6(2), Suppl. 1, 2020, Article ID: 2009045, DOI: 10.18416/IJMPI.2020.2009045