International Journal on Magnetic Particle Imaging IJMPI
Vol. 8 No. 1 Suppl 1 (2022): Int J Mag Part Imag
https://doi.org/10.18416/IJMPI.2022.2203061
Comparison of Reconstruction Methods for Measured FFL Data
Main Article Content
Copyright (c) 2022 Jonas Schumacher, Mandy Ahlborg, Eric Aderhold, Pascal Stagge, Thorsten M. Buzug, Matthias Gräser
This work is licensed under a Creative Commons Attribution 4.0 International License.
Abstract
System matrix and x-space resonstruction are two of the main approaches for image reconstruction of field free line magnetic particle imaging. A comparative studiy of both options is performed on data from a phantom
measurement in a permanent magnet-based scanner system. The system matrix reconstruction is performed in a hybrid fashion with data obtained in a spectrometer. This data is also used to obtain the relaxation time and particle diameter of the particles used. A deconvolution can then be used to enhance the image quality of the x-space approach which is compared to the system matrix reconstruction. A slight blurring can be depicted in the x-space reconstruction but overall a good agreement between both approaches is reached with the Structural Similarity Index yielding a value of 0.79.