International Journal on Magnetic Particle Imaging IJMPI
Vol. 9 No. 1 Suppl 1 (2023): Int J Mag Part Imag
https://doi.org/10.18416/IJMPI.2023.2303007
MPI Reconstruction Based on System Matrix using a Field Free Line
Main Article Content
This work is licensed under a Creative Commons Attribution 4.0 International License.
Abstract
Magnetic particle imaging (MPI) is a new imaging modality which draws much attention due to its high sensitivity and spatial resolution. The spatial information encoded using a field free line (FFL) exhibit better sensitivity and signal-to-noise ratio compared to field free point (FFP) scanning. In this study, we investigate different reconstruction methods for MPI image based on system matrix and FFL scanning using simulation data. Our results show that existing reconstructed methods are influenced by particle size, noise, iteration coefficient, and trajectory density. We evaluate the weighting iterative method and the results are shown with a phantom. This work bridges the gap between simulation and measurement experimental work by demonstrating the feasibility of reconstructing 2D images using a simulated MPI system matrix and FFL.